Innovatus Imaging repairs about 45 Philips X5-1 ultrasound probes EVERY month. And, that’s only client probes. Add about another third for refreshing those in our finished-goods loaner inventory. That’s a lot of X5-1s! Each probe (client or Innovatus owned) undergoes an incoming assessment, multiple in-process performance checks, and a final outgoing quality inspection. Doing the math, our quality team performs about 2,000 evaluations per year on JUST Philips X5-1s. We understand X5-1 failures.
Wiring Failures
The Philips X5-1 exhibits failures similar to other transducers…lens damage, strain relief damage, cable jacket damage, wiring intermittencies, connector damage, and random electronic failures. Traditionally, the primary mode of failure on almost all cardiac probes was a breakdown of the individual micro-coaxial wires within the wiring harness due to the way echo studies are performed.
Years of Philips X5-1 failure analysis data shows that wiring failures are still the most frequently encountered failure.
Cable failures and wiring intermittencies typically present as INTERMITTENT noise artifact (static) in CW Doppler mode or INTERMITTENT color streaking in color Doppler mode. Below is a procedure that you can use to interrogate your system cable in the field.
Begin troubleshooting by potentially eliminating or identifying if the root cause is cable related.
- Firmly connect the X5-1 to the scanner and initialize using a factory preset, such as Adult Echo. Be sure that all water and gel have been removed from the lens of the probe.
- Place the scanner in CW Doppler mode and move the sample volume cursor slightly left or right of center (Avoid a straight vertical/centered line).
- Adjust the Doppler gain so there is very minimal or no background noise in the spectral area.
- Flex each strain relief and the length of the cable while observing the scrolling spectral Doppler area and listening to the Doppler output.
- Move the sample volume cursor to multiple locations within the image and retest.
When testing factory new and slightly used probes, results will present audibly as, what sounds like, slight rubbing and visually as minor gray spikes, peaks and valleys in the spectral area. If the wires are beginning to fail (or have failed) hard/heavy static will be heard, and very pronounced white spikes will be visualized. The amount of acceptable static is highly subjective and can be interpreted differently by different people. Typically, more than one wire breaks down at a time and failures will be identified at multiple locations throughout the image. If constant static is encountered at some point, the point of failure may not be cable related.
Physical Damage
The least common problems with the Philips X5-1 are those which are physical, such a only lens damage, only strain relief damage, or only connector damage. Although many X5-1’s present with lens and strain relief damage, their main, underlying, failure is wiring-related. The 7% includes X5-1s with ONLY physical failures.
Array Failures and No Trouble Found
Understand X5-1 failures: Cable/wiring failures are still common on the X5-1, but all the technological advancements in this probe have presented new modes of failure, never seen before. Historically, the acoustic array on cardiac probes has had a very robust design. Very few (< 5%) actually experienced an array failure unless the victim of some significant trauma. The array, or “crystals” on the X5-1 is mounted to an ASIC (Application Specific Integrated Circuit). Its a revolutionary design, far beyond that of traditional cardiac probes, such as the Philips S5-1, GE M5S-D, or Siemens 4V1c and other non-3D volumetric probes. It’s just one of several reasons that the X5-1 costs 2-3-times as much as its standard counterpart, the S5-1.
Another very common mode of failure with this probe model is CONSTANT noise (or static) in CW Doppler mode at only specific locations within the image. The problem may or may not actually be repeatable, which makes it extremely frustrating for echo-techs as well as service personnel. A significant percentage (12%) of the X5-1’s exhibiting this phenomenon may actually have no problem whatsoever, however; it is likely that it is the most frequent reason for which the probe is replaced.
The engineers at our Center of Excellence have developed custom fixtures and new test methodologies to isolate and interrogate various electronics within these probes and after thorough research, we’ve identified that IF a poor, intermittent, compromised or questionable connection results on one or more of the 12 control lines within the X5-1, then constant static in CW Doppler mode may be experienced. Ultimately, ANYTHING that compromises the connection of one or more of the control lines leading to the array may result in, what seems to be, a hard probe failure. Obviously, a failure in the array/ASIC could also lead to this symptom.
Understand X5-1 failures: ANYTHING that compromises the quality of the connection or conductivity of one or more of the 12 control lines leading to the ASIC may result in constant static.
Before sending in an X5-1 for CONSTANT noise in CW Doppler mode, follow the steps below
- Begin troubleshooting by thoroughly inspecting the pin bank on the reported probe. Then inspect ALL ports on ALL scanners on which the probe MAY be used. Under magnification, inspect EACH pin. Look for deformations, compressed areas, oxidation and corrosion. The Compact Extreme (CX/Epiq) style connectors can be challenging to fully visualize. Use items such as headset magnifiers and dental mirrors to assist. Defects can be easily missed. It’s also possible that a single bent pin on a single probe has compromised the connectors across multiple ports on multiple scanners. The damage to one port, then affects the next probe connected. So on and so forth.
- Consider that the spring tension of the pins on either the single X5-1 or the scanner(s) has weakened or deteriorated to a point where it affects the Philips X5-1.
- Don’t assume that because other probe models don’t experience this scenario that the problem is not scanner related. So far, we’ve identified that this failure is isolated to the Philips X5-1 ONLY.
- Also, don’t assume that replacing the probe will solve the issue. It may, but may only for the short-term. The spring tension of the pins on new probes may be better than on used/older probes. Using a new probe on a scanner that has worn/compromised connectors can result in the same problem arising later.
When complaints of constant static arise, actively work with your sonographers. When the issue is encountered, or when troubleshooting, work together.
- Have the sonographer disconnect and reconnect the probe. This may solve a temporary seating/connectivity issue
- If the problem persists, disconnect the probe and reconnect it to a DIFFERENT port. Spring tension can wear and decrease. Oxidation can build up on probes/scanner ports over time, even if the probe is never disconnected.
- Finally, test the affected probe on another scanner. As with most performance testing within ultrasound, consistency is key. Test the probe using the IDENTICAL settings with the Doppler sample volume in the IDENTICAL location. To help, capture photos of the scanner settings and sample volume locations using a mobile device.
On the surface, replacing a probe may seem like the easiest solution to troubleshoot or solve this particular failure. With today’s more advanced probe designs, you may just need better tools in your toolbox. Make sure to partner with a provider that has advanced as much as the products you support.
Access this material in a downloadable version below